品牌AIRTIMES/中航时代

ZST-121橡胶塑料电阻率测定仪(体积表面电阻率测试仪)


符合标准: 
GB1410《固体电工绝缘材料绝缘电阻、体积电阻系数和表面电阻试验方法》 
ASTM D257《绝缘材料的直流电阻或电导试验方法》 

一、概述 
    本仪器既可测量高电阻,又可测微电流。采用了美国In公司的大规模集成电路,使仪器体积小、重量轻、准确度高。以双3.1/2 位数字直接显示电阻的高阻计和电流。量限从1×104Ω ~1×1018 Ω,是目前国内测量范围zui宽,准确度zui高的数字超高阻测量仪。电流测量范围为2×10-4 ~1×10-16A。机内测试电压为10/50/100/250/500/1000V任意可调。本仪器具有精度高、显示迅速、性好稳定、读数方便, 适用于防静电产品如防静电鞋、防静电塑料橡胶制品、计算机房防静电活动地板等电阻值的检验以及绝缘材料和电子电器产品的绝缘电阻测量。本仪器除能测电阻外,还能直接测量微弱电流。

三、技术指标 
1、电阻测量范围: 1×104Ω ~1×1018Ω。 
2、电流测量范围为: 2×10-4A~1×10-16A 
3、双表头显示: 3.1/2位LED显示 
4、内置测试电压:10V、50V、100V、250、500、1000V 
5、基本准确度:1% (*注) 
6、使用环境: 温度:0℃~40℃,相对湿度<80% 
7、机内测试电压: 10/50/100/250/500/1000V 任意切换 
8、供电形式: AC 220V,50HZ,功耗约5W 
9、仪器尺寸: 285mm× 245mm× 120 mm 
10、质量: 约2.5KG 

四、工作原理 
    根据欧姆定律,被测电阻Rx等于施加电压V除以通过的电流I。传统的高阻计的工作原理是测量电压V固定,通过测量流过取样电阻的电流I来得到电阻值。从欧姆定律可以看出,由于电流I是与电阻成反比,而不是成正比,所以电阻的显示值是非线性的,即电阻无穷大时,电流为零,即表头的零位处是∞,其附近的刻度非常密,分辨率很低。整个刻度是非线性的。又由于测量不同的电阻时,其电压V也会有些变化,所以普通的高阻计是精度差、分辨率低。 
    本仪器是同时测出电阻两端的电压V和流过电阻的电流I,通过内部的大规模集成电路完成电压除以电流的计算,然后把所得到的结果经过A/D转换后以数字显示出电阻值,
即便是电阻两端的电压V和流过电阻的电流I是同时变化,其显示的电阻值不象普通高阻计那样因被测电压V的变化或电流I的变化而变,所以,即使测量电压、被测量电阻、电源电压等发生变化对其结果影响不大,其测量精度很高(),从理论上讲其误差可以做到零,而实际误差可以做到千分之几或万分之几。 

典型应用 
1.测量防静电鞋、导电鞋的电阻值 
2、测量防静电材料的电阻及电阻率 
3、测量计算机房用活动地板的系统电阻值 
4、测量绝缘材料电阻(率) 
5、光电二极管暗电流测量 
6、物理,光学和材料研究 

标准配置: 
1、测试仪器:1台 
2、电源线:1条 
3、测量线:3根(屏蔽线、测试接线、接地线) 
4、使用说明书:1份

橡胶塑料电阻率测定仪使用方法 
6.1接好电源线 
确保电源为220VAC/50Hz 
6.2接通电源 
   将电流电阻量程置于 104 档,电压量程置于10V,然后开机。
6.3调零 
在“Rx"两端开路的情况下,调零使电流表的显示为0000 .注意:在“Rx"两端不开路,如接在电阻箱或被测量物体上时调零后测量会产生很大的误差。一般一次调零后在测试过程中不需再调零。 完毕后关机。
6.4连接线路 
    接好测试线,将测试线将主机与屏蔽箱连接好,测体积电阻时测试按钮拨到Rv边,测表面电阻时测试按钮拨到Rs边。然后开机。
6.5选择合适的测量电压 
★电压选择开关在后面板,注意,在测试过程中不要随意改动测量电压,可能因电压的过高或电流过大损坏被测试器件或测试仪器; 
6.6测试 
测量时从低档位逐渐拔往高档,每拨一次稍停留1~2秒以使观察显示数字, 当被测电阻大于仪器测量量程时,电阻表显示“1",此时应继续将仪器拨到量程更高的位置,当测量仪器有显示值时应停下,当前的数字乘以档次即是被测电阻值。当有显示数字时不要再往更高次档拨,否测仪器会过量程,机内保护电路开始工作,仪器测量准确度会下降。 

6.7 测试完毕将电阻电流量程拔至“104 "档,电压量程调至10V后关闭电源
★每测量一次均应将量程开关拨回到104“调零"档的量程位置以免开机或测量端短路时而损坏仪器。
6.8 测量电流及1015Ω以上超高电阻的测量应用测量电流后用欧姆定律以电压除以电流计算电阻的方法,详见8.5节内容。
6.9 体积电阻和表面电阻转换
★在测试过程中,使用屏蔽箱在进行体积电阻和表面电阻转换时,必须把电源关闭后进行档位转换,否则会导致电压冲击到主机无法显示或损坏。

Meet the standard:

GB1410 "solid electrical insulation materials insulation resistance, volume resistivity and surfaceresistance test method"

DC resistance or conductance of insulating materials test method "ASTM D257"

One, overview

This instrument can measure high resistance, and micrometer current. Using large scale integrated circuit American In company, so that the instrument has the advantages of small volume, light weight, high accuracy. High resistance meter and current resistance directly by double 3.1/2 digit display. Range from 1 * 104 to 1 * 1018 Omega Omega, is currently the mostwide measuring range, digital with the highest accuracy of ultra high resistance measuring instrument. Current measuring range of 2 * 10-4 to 1 * 10-16A. The built-in test voltage of 10/50/100/250/500/1000V adjustable. The instrument has high precision, good stability, showing rapid, convenient reading, suitable for measuring insulation resistance in anti-static productssuch as anti-static shoes, anti-static plastic and rubber products, anti static floor computer roominspection such as value of resistance and insulation materials and electronic products. In addition to the instruments can measure resistance, but also the direct measurement of weakcurrent.

Three, technical indicators

1, resistance measuring range: 1 * 104 to 1 * 1018 Omega omega.

2, the current measuring range: 2 * 10-4A to 1 * 10-16A

3, double headers: 3.1/2 LED display

4, the built-in test voltage: 10V, 50V, 100V, 250, 500, 1000V

5, basic accuracy: 1% (note)

6, the use of the environment: temperature: 0 to 40 DEG C, relative humidity <80%

7, the built-in test voltage: 10/50/100/250/500/1000V arbitrary switching

8 power supply forms: AC, 220V, 50HZ, power consumption is about 5W

9, the instrument size: 285mm x 245mm x 120 mm

10, quality: about 2.5KG

Four, the principle of work

According to Ohm's law, the resistance being measured Rx is equal to the applied voltagedivided by current V by I. The working principle of the high resistance meter is the traditionalmeasuring voltage V fixed, by measuring the current flowing through the sampling resistor I toobtain the resistance value. From Ohm's law can be seen, because the current I is inversely proportional to the resistance, but not proportional, so is nonlinear resistive display value, i.e.,infinite resistance, current is zero, i.e. zero at the header is infinite, scale near the very dense,with a low resolution. The scale is nonlinear. Because the resistance measurement of differentvoltage, the V will also be some changes, so the high resistance of ordinary gauge is poor accuracy and low resolution.

This instrument is the current flowing through the resistor V and a voltage across the resistor Imeasured at the same time, the large scale integrated circuit internal to complete the calculation of voltage divided by current, then put the results obtained after A/D conversion to digital displayresistance, even current voltage V and the flow resistance at both ends of the I is changed at the same time. The resistance of the display value is not like an ordinary high resistance meterthat due to changes in the voltage to be measured changes in V or I current and change, so,even if the measure voltage, measure resistance, power supply voltage change has little influence on the results, the measuring accuracy is very high (patent), theoretically the error can do zero, while the actual error can do a few thousandths or very few.

Typical applications

1 resistance measurement of anti-static shoes, conductive shoes value

2, measurement of anti electrostatic material resistance and resistivity

3, measurement of computer room with movable floor system resistance value

Insulation resistance 4, measurement (rate)

5, photoelectric diode dark current measurement

6, physics, optics and material research

Standard configuration:

1, test equipment: 1 sets

2, the power line: 1

3, measuring line: 3 (shielding wire, test wiring, grounding wire)

4, instructions for use: 1

Six using the method

6.1 connect the power line

Make sure the power is 220VAC/50Hz

6.2 connect the power supply

The current resistance range in 104 files, voltage range in 10V, and then boot.

6.3 zero

In the "Rx" at both ends open under the condition of zero to display the current table 0 note: do not open in the "Rx" ends, such as connected to the resistance box or an object to be measuredon the tune measurement of zero after the great error will occur. A general tone in the testing process does not need to be zero zero. After shutdown.

The 6.4 connection line

Connect the test line and testing line will host and shielding box is connected well, the test button dial to Rv edge of measuring volume resistance, surface resistance measurement when the test button dial to Rs side. And then boot.

6.5 select the appropriate measuring voltage

* the voltage selector switch on the rear panel, attention, in the test process, do not arbitrarily change the voltage measurements, probably due to the high voltage or current is too large and damage to the device or testing instrument is tested;

The 6.6 test

When measuring low level to high level gradually from pull out, each dial again slightly stay 1 to2 seconds to make the observation of digital display, when the measured resistance greater than the instrument measurement range, resistance meter shows "1", this time should continue to instrument dial to range higher position, when measuring instruments have displayed valueshould be stop the current number multiplied by the grade is measured resistance values. When the display numbers do not go down to the higher gear shifting, no measuring instruments willrange, machine protection circuit starts to work, the instrument measuring accuracy will decline.

The 6.7 test is completed the resistance current range pull to "104", the voltage range to 10Vafter power off

* each measuring a shall measuring range switch back to 104 "range position zero" file so as not to boot or measuring terminal short circuit and damage the instrument.

The 6.8 methods above to measure current and 1015 ohm measurement application of measuring current and very high resistance after using Ohm's law to the voltage divided by thecurrent calculation of resistance, described in Section 8.5 content.

6.9 the volume resistance and surface resistance conversion

* in the testing process,

产品咨询

留言框

  • 产品:

  • 您的单位:

  • 您的姓名:

  • 联系电话:

  • 常用邮箱:

  • 省份:

  • 详细地址:

  • 补充说明:

  • 验证码:

    请输入计算结果(填写阿拉伯数字),如:三加四=7
北京中航时代仪器设备有限公司
  • 联系人:石磊
  • 地址:北京市房山区经济技术开发区1号
  • 邮箱:zhsdyq@163.com
  • 传真:86-010-80224846
关注我们

欢迎您关注我们的微信公众号了解更多信息

扫一扫
关注我们
版权所有 © 2024 北京中航时代仪器设备有限公司 All Rights Reserved    备案号:京ICP备14029093号-1    sitemap.xml
管理登陆    技术支持:化工仪器网